Teams in Hefei, Jinan and Nanjing compared two systems that assist surgeons during prostate removal. The study by Cong Huang and colleagues, published online on 12 November, enrolled 80 patients with localised prostate cancer, most of them at low or intermediate risk.

Between May 2023 and January 2024, patients were randomly assigned to surgery with the Edge MP1000 or the da Vinci Xi, 40 to each group. Every planned procedure could be completed without switching techniques. The observed difference on this endpoint was zero. The 95% confidence interval ran from −8.8 to +8.8 percentage points, with its lower bound above the trial’s prespecified −20-point non-inferiority margin.
This result describes the surgical teams’ ability to complete the procedure. The researchers had 12-week follow-up data for 73 patients. A comparison of long-term cancer control or rare adverse events would require a larger study and longer follow-up.